Meeting FDA cybersecurity guidance for AI-enabled device software

The FDA's January 2025 AI-Enabled Device guidance identifies seven AI-specific cyber threats. Only one of them, model evasion, lives at runtime on the device. The other six attack the pipeline: training data, model registries, evaluation logic, and deployment.

7 FDA-named AI threats. Only 1 lives at runtime.Data poisoningPIPELINEModel inversionPIPELINEData leakagePIPELINEOverfittingPIPELINEModel biasPIPELINEPerformance driftPIPELINEModel evasionRUNTIME
Six of the seven FDA-named AI threats attack the pipeline, not the deployed model.

Most testing programs today only cover the deployed model. That means they are addressing one-seventh of what the FDA is explicitly asking about. Data poisoning, model inversion and stealing, data leakage, overfitting, model bias, and performance drift all sit upstream of the running model.

Test everything that touches the model

Our approach expands the digital twin to include the full AI pipeline as an associated system, then structures testing to cover the whole lifecycle. Runtime testing stays focused and purpose-built. Every test case maps back to a specific FDA-identified threat, executed against real pipeline components, producing defensible evidence for premarket and postmarket.

This is not "test the model." It is "test everything that touches the model."
← All intelligence
AI inside the device

The model ships
without its pipeline.

A model is trained off the device, and so is most of the evidence that it is safe. The version that ships is the last frame of a long process, and cybersecurity that stops at that frame misses where the risk was built.

THE AI LIFECYCLE, AND WHERE THE SEVEN THREATS LIVETRAINING DATAsourced, labeled, governedData poisoningModel biasData leakageMODEL TRAININGfit, tuned, evaluatedOverfittingPerformance driftMODEL REGISTRYversioned, stored, servedModel inversionModel stealingData leakageDEPLOYED DEVICEthe model at runtimeModel evasionTESTED BY MOST SUBMISSIONSmodel evasion, 1 of 7COVERED BY THE JANUARY 2025 GUIDANCEall seven threats, across the whole pipeline, 7 of 7

Well over a thousand AI-enabled devices sit on the FDA's public list, and most premarket submissions scoped cybersecurity to the deployed model alone. The January 2025 draft says the scope is the whole lifecycle.

FDA draft, Section XII

Seven AI threats.
Most programs test one.

Section XII of the January 2025 draft names seven cybersecurity threats specific to AI-enabled devices. Six of them attack the data and the pipeline that produce the model. Only one, model evasion, happens at runtime.

Data poisoning

Inauthentic or maliciously modified data injected into the training set.

Lives in: Training data

Model inversion or stealing

Forged or altered inputs used to infer training data from, or replicate, the model.

Lives in: Model and inference

Model evasion

Inputs crafted to deceive the model as it runs on the device.

Lives in: Deployed device

Data leakage

Access to sensitive training or inference data.

Lives in: Pipeline and deployment

Overfitting

AI components pushed with adversarial input so a narrowly fit model returns skewed outcomes.

Lives in: Training and evaluation

Model bias

Training data manipulated to introduce or amplify bias.

Lives in: Training data

Performance drift

The underlying data distribution shifted so the model degrades over time.

Lives in: Postmarket
A program scoped to the deployed model answers one of the seven. The guidance asks about all seven, in the premarket submission and across the postmarket lifecycle. Sponsors of an AI-enabled device that meets the definition of a cyber device under section 524B must document these risks in the marketing submission, and the draft supplements the FDA's 2023 premarket cybersecurity guidance rather than replacing it.
Timing

You cannot test poisoning
on a frozen model.

Most of the seven threats are decided months before a device ships. Poisoning resistance, bias, and generalization are properties of how the model was trained, not switches you can flip at submission time.

The standards landscape

The rules are arriving,
and they are not harmonized.

The FDA draft does not stand alone. A wave of AI and data standards has landed recently, each reaching a US submission from a different angle, none of them yet aligned with the others.

Featured · published 2026

IEC PAS 63621: data management for AI-enabled devices

It governs data across its whole lifecycle: provenance, integrity, and confidentiality, with protection against tampering and loss. It introduces Data of Unknown Provenance, or DOUP, for data whose source, date, or validity cannot be confirmed, and integrity insurance, the set of measures that keep data accurate and unaltered end to end. Six of the FDA's seven threats attack exactly this layer, which is why a data-management standard is now a cybersecurity document.

ISO/IEC 42001

Management system for artificial intelligence: the governance wrapper around how an organization builds and runs AI.

ISO/IEC 23894

Risk management guidance for AI, the AI-specific companion to enterprise risk practice.

ISO/IEC 27559

Privacy-enhancing data de-identification, referenced inside IEC PAS 63621 for the confidentiality of datasets.

AAMI CR34971

Guidance on applying ISO 14971 risk management to machine learning in medical devices.

NIST AI RMF

The US framework for identifying and managing AI risk across the lifecycle.

IMDRF ML principles

International good machine learning practice, the harmonization effort regulators are converging toward.

None are harmonized. All of them still reach you.

What good looks like

Evidence mapped
to every named threat.

A defensible AI cybersecurity package does one thing well. It maps each test and each control back to a specific named threat, across the pipeline and the deployed model, so a reviewer can follow it.

Treat the pipeline as in scope

Model the training data, the registries, the evaluation logic, and deployment as associated systems inside the device, not as somebody else's problem. That is what lets testing cover the lifecycle instead of the last version.

One threat, one line of evidence

Every test case and result points at the specific FDA-named threat it answers, premarket and postmarket. Coverage becomes a list a reviewer can read, not a claim they have to take on faith.

Runtime testing stays

Model evasion is still a runtime threat on the real device, so that testing keeps its place. The guidance widens the scope, it does not shift it. The instruction is not to test the model, it is to test everything that touches the model.

Sources: FDA draft guidance, Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendations, January 2025, Section XII. IEC PAS 63621:2026, Artificial intelligence enabled medical devices, Data management.
Get started

See your device through ELTON.

Start with one device. We build the twin from documentation your quality system already produces, run AI discovery remotely, and show you the graph: the handful to fix, and the evidence for everything else.

Proof Over Probability

The AI testing newsletter.

One issue a month on AI, exploitability, and FDA cybersecurity review. No spam, unsubscribe anytime.

Questions

Common questions about the FDA AI-enabled device guidance.

What AI cybersecurity threats does the FDA identify for AI-enabled devices?

The FDA's January 2025 AI-Enabled Device guidance names seven AI-specific cyber threats: data poisoning, model inversion and stealing, data leakage, overfitting, model bias, performance drift, and model evasion. Only model evasion lives at runtime on the device. The other six attack the pipeline that produces the model.

Is testing the deployed model enough to meet the FDA AI guidance?

No. A program scoped to the deployed model covers one of the seven threats the FDA names, model evasion. The other six sit upstream in training data, model registries, evaluation logic and deployment, which means runtime-only testing addresses roughly one seventh of what the guidance asks about.

What counts as the AI pipeline for a medical device?

The AI pipeline is everything that touches the model before it runs: training data, model registries, evaluation logic and deployment. Treating that pipeline as an associated system inside the device model is what lets testing cover the whole lifecycle instead of stopping at the version that ships.

How do you produce defensible evidence for AI-specific threats?

Map every test case back to a specific FDA-identified threat and execute it against real pipeline components. That mapping is what turns test activity into evidence a reviewer can follow for premarket and postmarket, because each result points at the named threat it answers.

Does the AI guidance replace runtime security testing?

No. Runtime testing stays in scope and stays focused, because model evasion is a runtime threat. The guidance widens the scope rather than shifting it. The instruction is not to test the model. It is to test everything that touches the model.

What does IEC PAS 63621 require for AI training data?

IEC PAS 63621, published in 2026, is a data-management specification for AI-enabled medical devices. It asks manufacturers to govern data across its lifecycle: establish provenance, maintain integrity and confidentiality, control access, and protect against tampering and loss. It introduces Data of Unknown Provenance, or DOUP, for data whose source cannot be confirmed, and integrity insurance for keeping data accurate end to end. Because six of the FDA's seven threats attack the data layer, it reads as a cybersecurity document as much as a quality one.

The FDA guidance is still a draft. Do I need to act now?

The guidance is draft, but it reflects what reviewers are already asking about AI-enabled devices, and the underlying obligation is law: a device that meets the section 524B definition of a cyber device must address cybersecurity in its submission. Published standards like IEC PAS 63621 are already in force, and premarket evidence for AI properties has to be captured during development, not reconstructed later. Waiting for the final guidance does not buy back the training-time evidence you would have needed to collect.

Which standards apply beyond the FDA guidance?

Several, and they are not yet harmonized. IEC PAS 63621 covers AI data management. ISO/IEC 42001 and ISO/IEC 23894 cover AI management systems and AI risk management. ISO/IEC 27559 covers dataset de-identification. AAMI CR34971 applies ISO 14971 risk management to machine learning, and the NIST AI RMF and IMDRF good machine learning practice add US and international framing. Each reaches a US submission from a different angle, so a program built around the FDA's seven threats gives you a spine the other standards map onto.

Exploitability management for medical devices. FDA §524B methodologyExploitability proven at runtime95% faster than legacy testing Book a Demo →
Platform
OverviewAvoid FDA DeficienciesAvoid Consulting FeesDigital Twin TraceabilityAI MedDevice PentestingExploitability VerificationVulnerability ChainingRemediation OptimizationRemote TestLink™Incident ResponseAutomated VEX & MetricsCVSSv4 Migration
Solutions
EnterpriseStartups / SMBs Postmarket SurveillanceIncident ResponseSecurity EngineeringRegulatory AffairsFDA §524BEU MDR/CRAEU REDNIS2IMDRF N60 / N73Japan MHLW
Why ELTON
Proof over ProbabilityExploitability VerificationFDA MDDTCVSSv4 MigrationQMSR Audit ComplianceAI-Enabled Device Testing One SolutionSubscription TestingAI-NativeELTON vs. Legacy TestingThreat-Led AI PentestingCredentialsDevice ModalitiesPricing
Resources
FDA Deficiency ListFDA Testing RequirementsFDA Cyber SOPs & TemplatesRemediation LibraryRegulatory GuidesWebinarsAI NewsletterThe End of Legacy TestingThe AI Vulnerability ExplosionAI Inside the ProductCybersecurity TestingSecurity AdvisoriesWhitepapersIntelligence & Blog
Company
AboutLeadershipCareersPartnershipsData SecurityContact Meet ELTON →